Filter Media Test System

The Test System for characterisation of cleanable filter media acc. VDI 3926-2. Testing of filter media is important for the development and optimisation of filter materials as ...

Introduction

The Test System for characterisation of cleanable filter media acc. VDI 3926-2. Testing of filter media is important for the development and optimisation of filter materials as well as for quality assurance during the production process.

Most interesting parameters for filter media characterisation are:

– differential pressure drop

– gravimetric separation efficiency

– particle size dependent filtration efficiency

– dust loading capacity

For these tasks We developed the Filter Media Test System which facilitates filter tests in accordance with guideline VDI 3926 exemplary embodiment 2 and ASTM D6830-02, but can also be used for other tests. Separation efficiency and dust loading tests on basis of the general requirements of ISO 16890, EN 779, DIN 71460-1, ISO/TS 11155-1 and ISO 5011 can be performed in small scale. Due to its compact and lightweight design this test rig can be used in the laboratory.

Applications

filter tests for cleanable filter according to VDI 3926-2

determination of fractional separation efficiency, dust loading capacity and differential pressure drop of filter mediafilter tests based on other standards (air cleaning filters ISO 16890, EN 779; cabin air filters ISO 11155-1)

development and optimisation of filter mediaquality assurance measurements

Standards

ISO 12103-1

 

ISO 5011

VDI 3926-2

Benefits

standardised filter testing with alumina oxide (e.g.: Pural NF)

automated testing procedure over 14 hours without intervention by operator

use of different aerosols possible (solid and droplet aerosols)

flexible design enables fast modification and future addons

operatir-friendly control software

Technical data

Parameter Unit Value
Test flow rate m³/h 2 … 64
Measuring range, differential pressure Pa
Duration of pressure pulse ms 50 … 150
Pressure tank capacity bar 2 … 6
Test specimen dimensions mm Diameter ≤ 150
Power supply 3 × 400 VAC, 16 A
Operating medium (aerosol substance) Pural-NF, Pural-SB, DEHS
Sensors (others) Temperature, relative humidity
System dimensions (W × H × D) mm 2800 × 1500 × 800
Weight kg approx. 120

Accessories

Atomizer Aerosol Generator

Solid materials disperser

Aerosol Dilution System

 

Aerosol Spectrometer

Inquiry