Battery Isothermal Calorimeter
Category:Battery performance tester
Introduction
Product Description
Battery Isothermal Calorimeter is a professional instrument developed based on the principle of power compensation isothermal calorimetry for testing the thermal characteristics of various types of lithium batteries, capable of realizing the measurement of the thermal characteristics of lithium batteries’ charging and discharging as well as the thermal and physical parameters, providing accurate, stable and reliable basic thermal data for the thermal simulation of batteries, the optimization of the design of the thermal management system, as well as the assessment of the thermal safety performance of batteries.
Product Features
1. In an isothermal environment, it is compatible with the power compensation method and the heat – flow method to measure the heat generated during battery charging and discharging. This not only ensures the accuracy of thermal characteristic measurement for relatively large – sized batteries but also improves the sensitivity of heat measurement for relatively small – sized batteries.
2. It has an instrument calibration mode, which can calibrate the measurement accuracy of charging and discharging thermal behavior parameters under different conditions.
3. It is equipped with a heat capacity measurement mode. By using the comparative method to measure heat capacity, it reduces the operational requirements for heat capacity testing and can accurately measure the heat capacity of batteries at different temperatures.
4. The integrated battery charging and discharging module enables the switching of charging and discharging modes, the setting of constant – current/constant – voltage charging modes, the setting of charging/discharging currents, and the real – time calculation of battery power.
5. Network communication is adopted to achieve remote operation of the instrument and remote data transmission, ensuring the safety of testers.
6. It can simultaneously record voltage, current, and temperature data during the battery charging and discharging process.
7. An optional nitrogen purging module can be selected to ensure the accuracy of low – temperature measurement and effectively suppress the problem of low – temperature condensation.
Technical Specifications Parameters
Parameter | Value |
---|---|
Calorimeter Body Operating Environment | (5~40)°C, <85%RH |
Oil Bath Temperature Control Range | (-40~100)ºC |
Experimental Modes | Isothermal Power Compensation Mode, Isothermal Heat Flow Mode, Specific Heat Capacitance Mode |
Temperature Stability | ±0.005°C |
Temperature Resolution | 0.001°C |
Maximum Cell Size | L345mm×W230mm×H100mm |
Maximum Compensated Power | 200W |
Calorimetric Sensitivity | 10mW (power compensation mode); 0.2mW (heat flow mode) |
Absorption and Expiration Enthalpy Measurement Accuracy | ±1% (power compensation mode); ±2% (heat flow mode) |
Baseline Noise | 10mW (Power Compensation Mode); 0.2mW (Heat Flow Mode) |
Heating Channels | Standard with 2 channels |
Temperature Measurement Channels | Standard with 8 channels |
Instrument Interfaces | RJ45 |
Adjustable Gas Flow Rate | (5~25)L/min |
Power Supplied Power Supply | AC220V/50Hz |
Maximum Power | 600W |
Oil Bath Setting Temperature Range | (-55~200 )°C |
Oil Bath Temperature Stability | ±0.01°C |
Oil Bath Display Resolution | 0.01°C |
Heating Power | 3kW |
Refrigeration Power | 1.5kW@20°C, 1.5kW@0°C, 1kW@-20°C, 0.3kW @-40°C |
Bath Openings/Size | 12×11/16cm |
Charging | 5L |
Instrument Interfaces | Serial port |
Calorimeter Body Operating Environment (repeated) | (5~40)°C, <85%RH |
Oil Bath Temperature Control Range (repeated) | (-40~100)ºC |