Mask BFE Testing Machine

Mask BFE Testing Machine is used to detect the filtration efficiency of masks, and also can be used for measurement and calibration departments, scientific research institutions...

Category:Mask Tester

Introduction

Mask BFE Testing Machine is used to detect the filtration efficiency of masks, and also can be used for measurement and calibration departments, scientific research institutions, mask production enterprises and other relevant departments.

Features:

1. Mask BFE Testing Machine: negative pressure test system, can protect the safety of the operator.

2. Peristaltic pump built in the negative pressure cabinet, A, B channel 28.3L/min.

3. The flow rate of the peristaltic pump can be set.

4. BFE Tester atomization effect is good.

Mask BFE Testing Machine test principle:

Bacterial filtration efficiency Tester consists of aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure cabinet, 28.3 l/min carbon sampler and other components.The entire detector is controlled by the console.The console adopts microcomputer control system to coordinate and control the work of aerosol generating system, transmission system, negative pressure cabinet and sampling system, and display the working status in real time.The entire measurement work is done automatically.

Applicable Standards

ASTM F2100 standard specification for performance of materials used in medical face masks

EN 14683 standard that medical face masks in Europe are required to comply with

ASTM F2101 standard test method for evaluating bacterial filtration efficiency (BFE) of medical mask materials using Staphylococcus aureus biological aerosol

YY0469-2011 Technical requirements for surgical mask

YY/T 0969-2013 Single-use medical face mask

Technical Specification:

Conforms to EN 14683 and ASTM F2101

A Route Sampling Flow: 28.3L/min, resolution 0.1. E±2.O% L/min

B Route Sampling Flow: 28.3L/min, resolution 0.1. E±2.O% L/min

Spray Flow: (O.l e?? 10)L/min, resolution 0.1. E±5.O% L/min

Peristaltic Pump Flow: (0.001 ? 3.O)mL/min, resolution 0.001. E±2.O%

Negative Pressure Chamber: (-90 ? -120)Pa, resolution 0.1. E±I.O%

Working Temperature: OOC to 500C

High Efficiency Particulate Air Filter Properties:

≥99.995%@0.3μm, ≥99.995%@0.12μm

Diameter of aerosol generator mass: Average diameter:

(3.0+0.3)fm; Geometric Standard Deviation<1.5pm

Double Routes 6-lever Andersen Sampling apparatus:

Level l: >7gm; Level Il: (4.7 to 7)gm; Level Ill: (3.3 to 4.7)gm;

Level IV: (2.1 to 3.3)gm; Level V: (1.1 to 2.1)gm;

Level VI: (0.6 to 1.1)gm

Total number of positive quality control sampler particles: (2200±500)cfu

Negative Chamber Flow Speed: 25m/min

Size of Aerosol Chamber: at least (600x80x3) mm

Power supply: 220 V AC, 60 1-Iz

Noise: 65db(A)

Characteristics of G299 Mask BFE Testing Machine:

1. Intelligent touch screen control panel

more efficient and reliable, programmable 10 inch color LCD touch, achieve control, data display and other functions

2. High precision

Two samplers simultaneously sampling, avoid the test error by intermediate link, improve the accuracy of test

3. Multi directional protection design

Equipped with current overload protection, leakage protection and negative pressure cabinet to ensure safety for operator

4. Stability customized parameters

Adopts imported accessories, features stable movements, accurate test data and good repeatability.

 