Test System for the characterisation of the air flow velocity profile of components
Category:Filter Cartridge Filtration Tester
Introduction
Test System for the characterisation of the air flow velocity profile of components (e.g. tubes, filters).The air flow velocities between 1-40 m/s and turbulences in the outlet of flowed through units under test can be measured.
Through this, a good position for flow velocity measuring devices (for example MAFs) in flowed through parts (for example motor inlet filters) can be found easier, taking into consideration a range of air flow rates. The data created by the flow scanner can also be used to check models for flow simulation.
Applications
development of filter housings and air-ducting elements in general
comparison of models for flow simulation with measured values
Benefits
wide range of air flow rates 20 … 900 m³/h (expendable up to 1200 m³/h)
up to 100 hPa of pressure drop over unit under test and periphery (connectors, hoses)
fast data acquisition (turbulences) and graphic display of data
Principle
Air is blown through the test specimen and the velocity field is measured on the downstream side with the aid of a hot-wire anemometer. This automatically scans a plane in the air outlet by means of a traversing device.
Technical data
| Parameter | Unit | Value |
|---|---|---|
| Test flow rate | m³/h | 20 … 900 |
| Air flow velocity | m/s | 1 … 40 |
| Test specimen dimensions | mm | max. 485 × 600 |
| Differential pressure of test specimen | Pa | – |
| System dimensions (W × H × D) | mm | 2500 × 2660 × 1300 |
| Weight | kg | approx. 550 |
Accessories
hot wire anemometer set
picoscope (PC-oscilloscope)