Filter Test System for gas adsorption
Category:Filter Cartridge Filtration Tester
Introduction
The Filter Test System for gas adsorption/desorption of air filters for passenger compartments (DIN 71460-2, ISO 11155-2). The Adsorption Test Stand for air filters for passenger compartments is used for quantification of adsorption capacity or desorption behaviour of filters.
The design of the Adsorption Test Stand, the test procedure and data acquisition have been based on the draft standard DIN 71460-2 and ISO 11155-2 standard for measuring the dynamic gas adsorption.
The test section and associated ductwork are fabricated from stainless steel. Different sizes of filter media samples and elements can be tested and are mounted in a special adapter. A window in front of the test section allows visual observation of the sample during the test. A pair of mass flow controllers is used for dosing gaseous components from gas cylinders and a new, sophisticated, dosing unit for liquid gas components has been developed specifically for this application. This is based on evaporation of a droplet spray atomized by means of a two-stream nozzle fed from a µl-dosing pump. The gas concentration can be measured downstream of the filter. An exhaust system bypassing the test channel facilitates maintaining a very stable gas dosing.
Test acc. to standards DIN 71460-1 and ISO/TS 11155-1
For filter tests both acc. to standards DIN 71460-1, ISO/TS 11155-1 and E DIN 71460-2, ISO 11155-2
Standards
ISO 11155-2
DIN 71460-2
ISO 10121-2
ISO 10121-1
Benefits
standard compliant testing of filter elements and filter mediaoperator is guided through test routines
modular system and instrumentation
versatile
Principle
Integral parts of test procedures
– Differential-Pressure measurement
– Adsortion tests according standard ISO 11155-2 (DIN 71460-2)
– Adsorption/Desorption measurement
– user-defined test procedures
Technical Data
| Parameter | Unit | Value |
|---|---|---|
| Test flow rate | m³/h | 70 … 700 |
| Measuring range, differential pressure | Pa | 0 … 1000 |
| Test gases | – | n-Butane 80 ppm (C₄H₁₀), Toluene 80 ppm (C₆H₅CH₃), Sulphur dioxide 30 ppm (SO₂), Ammonia 30 ppm (NH₃), Nitrogen (purging) |
| Power supply | – | 3 × 230 VAC, 16 A |
| Test specimen dimensions | mm | max. 300 × 600 |
| Gas analysers | – | Hydrocarbons with FID, Sulfur oxides with NDIR, Nitrogen oxides with CLD |
| Sensors (others) | – | Temperature, relative humidity, absolute pressure |
| System dimensions (W × H × D) | mm | 1800 × 2500 × 1100 |
Accessories
test channel and rig
gas dosing system
exhaust system
flowmeter
gas analyser
sensors
test stand control software