Filter Test System for gas adsorption

The Filter Test System for gas adsorption/desorption of air filters for passenger compartments (DIN 71460-2, ISO 11155-2). The Adsorption Test Stand for air filters for passenge...

Introduction

The Filter Test System for gas adsorption/desorption of air filters for passenger compartments (DIN 71460-2, ISO 11155-2). The Adsorption Test Stand for air filters for passenger compartments is used for quantification of adsorption capacity or desorption behaviour of filters.

The design of the Adsorption Test Stand, the test procedure and data acquisition have been based on the draft standard DIN 71460-2 and ISO 11155-2 standard for measuring the dynamic gas adsorption.

The test section and associated ductwork are fabricated from stainless steel. Different sizes of filter media samples and elements can be tested and are mounted in a special adapter. A window in front of the test section allows visual observation of the sample during the test. A pair of mass flow controllers is used for dosing gaseous components from gas cylinders and a new, sophisticated, dosing unit for liquid gas components has been developed specifically for this application. This is based on evaporation of a droplet spray atomized by means of a two-stream nozzle fed from a µl-dosing pump. The gas concentration can be measured downstream of the filter. An exhaust system bypassing the test channel facilitates maintaining a very stable gas dosing.

Test acc. to standards DIN 71460-1 and ISO/TS 11155-1

For filter tests both acc. to standards DIN 71460-1, ISO/TS 11155-1 and E DIN 71460-2, ISO 11155-2

Standards

ISO 11155-2

DIN 71460-2

ISO 10121-2

ISO 10121-1

Benefits

standard compliant testing of filter elements and filter mediaoperator is guided through test routines

modular system and instrumentation

versatile

Principle

Integral parts of test procedures

– Differential-Pressure measurement

– Adsortion tests according standard ISO 11155-2 (DIN 71460-2)

– Adsorption/Desorption measurement

– user-defined test procedures

Technical Data

Parameter Unit Value
Test flow rate m³/h 70 … 700
Measuring range, differential pressure Pa 0 … 1000
Test gases n-Butane 80 ppm (C₄H₁₀), Toluene 80 ppm (C₆H₅CH₃), Sulphur dioxide 30 ppm (SO₂), Ammonia 30 ppm (NH₃), Nitrogen (purging)
Power supply 3 × 230 VAC, 16 A
Test specimen dimensions mm max. 300 × 600
Gas analysers Hydrocarbons with FID, Sulfur oxides with NDIR, Nitrogen oxides with CLD
Sensors (others) Temperature, relative humidity, absolute pressure
System dimensions (W × H × D) mm 1800 × 2500 × 1100

Accessories

test channel and rig

gas dosing system

exhaust system

flowmeter

gas analyser

sensors

test stand control software