Filter test system for general air filters

Filter test system for general air filters according to ISO 16890, EN 779, ASHRAE 52.2 The Filter Test System has been developed especially for air filter testing according to s...

Introduction

Filter test system for general air filters according to ISO 16890, EN 779, ASHRAE 52.2

The Filter Test System has been developed especially for air filter testing according to standards ISO 16890, EN 779 and ASHRAE 52.2. We can fall back to a long-standing expertise in the field of construction and installing of filter test systems.

In the filter test stand the filter samples can be easily installed into the rotatable and movable duct sections. A filtrated air flow is loaded with particles. In this way information about the efficiency of the filters will be obtained. With these results the filter samples can be rated and classified.

Applications

Testing of general air filters according to ISO 16890

Benefits

Measurements according to standards ISO 16890, EN 779:2012 and ASHRAE 52.2-2008

high grade of automation, mostly automated test runs with complete operator guiding through all tests, data handling and protocols

energy efficiency rating for filters according to EUROVENT

MERV-filter classification (Minimum Efficiency Reporting Value) for coarse dust filters and fine particle filters

rotatable duct sections

universal filter holding system

safety functions realized in a PLC

Technical data

Parameter Unit Value
Test flow rate m³/h 600 … 6000
Measuring range, differential pressure Pa max. 2000
Compressed air supply (operating gas/air) bar max. 6
Power supply 3 × 400 VAC, 125 A, 50 Hz
Operating medium (aerosol substances) DEHS, KCl, ISO 12103-1 A2 / A4

Accessories

TDC 584 – IPA Conditioning Cabinet for testing Air Filters acc. to ISO 16890-4

AFS 153 – Media tester